• page_banner01

Products

UP-6195 Customer Customization Floor Type Temperature&Humidity Chamber

GB2423/T5170/10586/10592, IEC68,GJB150,JIS C60068, ASTM D4714, CNS3625/12565/12566
The Test Chamber superior features and capabilities set it apart from traditional testing methods. Its precisely controlled temperature and humidity settings replicate the most stringent conditions to evaluate product performance and functionality in extreme environments.


Product Detail

Product Tags

Specification

Model

UP-6195

Inner size W×H×Dmm

500*400*400  500*600*500  600*850*800  100*100*80 100*100*100

Outside size W×H×Dmm

92*138*108 102*146*116 102*162*126 113*172*148 158*195*148 158*193*168

Inner chamber volumeV

80L  150L 225L  408L  800L  1000L

Temperature and humidity Controll range

-20℃150℃   -40℃150℃  -70℃150℃   RH20%-98%

Function

 

 

 

 

 

 

 

Temperature and humidty fluctuation

±0.5℃±2.5%RH

Temperature and humidity deviation

±0.5℃-±2℃±3%RH>75%RH; ±5%RH(≤75%RH)

Controller resolution

±0.3℃±2.5%RH

Controller brand

Korea TEMI880

Heating Time

RT-HT℃/min

100

150

 

40

50

Cooling Time

RT-LT℃/min

 

-20

-40

-60

-70

40

50

55

80

Power supply KW

4.0

6.0

6.0

6.0

Material

 

 

Inner material

SUS 304#  steel plate

Outside material

SUS 304#  steel plate(surface disposing)

Thermal insulation material

Thermostability high density chlorine formic acid foam insulator material

System wind pathround-robin mode

centrifugal fan-broadband enforce circulation

Refrigeration way

Single stage compression refrigeration

Refrigerator

whole sealing piston France Taikang brand Compressor

Freezing fluid

R4O4A or USA Dupont environmental protection refrigerantR23+R404

Condensation way

forced air cooling or water-cooling

Heater

Nichrome heating wire heating

Humidifier

Semi-enclosed steam humidification

Water-provide way

full-automatic  periodical feeding

Standard configuration

1 Observe windowdouble-deck cavity toughened glass,1  50mm test hole in the left,1  PL inner chamber lamp2clapboards, 1 bag of wet and dry bulb,3 fuses, 1 power line.

Safety device

Non- fuse-switchcompressor overload, refrigerant high low voltage,over-humidity and temperature protection,Protection switch,fuse stoppage warning system

Power

1φ 220V AC±10% 50/60Hz    3φ 380V AC±10% 50/60Hz