The three-in-one design makes the equipment easy to operate and space-saving.Users can do different testings of high temperature,low temperature and constant temperature humidity condition in every single testing area.
Every system is entirely independent each other,adopts 3 sets of refrigeration systems,3 sets of humidifying systems and 3 sets of controlling systems,so as to ensure stable and accurate controlling,and provide a longer service life.
Touch control & setting mode is totally controlled and locked up by automatic micro computer system with PID value automatic calculating ability.
Model No | UP6195A-72 | UP6195A-162 | |||||||
Inner chamber size(mm)W*H*D | 400×400×450 | 600×450×600 | |||||||
Exterior chamber size(mm)W*H*D | 1060×1760×780 | 1260×1910×830 | |||||||
Performance
|
Temperature range | -160℃,-150℃,-120℃,-100℃,-80℃,-70℃,-60℃,-40℃,-20℃,0℃~+150℃,200℃,250℃,300℃,400℃,500℃ | |||||||
Humidity range | 20%RH ~98%RH(10%RH ~98%RH or 5%RH ~98%RH) | ||||||||
Temp.&Humi fluctuation | ±0.2°C, ±0.5%RH | ||||||||
Temp.Humi.Uniformity | ±1.5°C; ±2.5%RH(RH≤75%),±4%RH(RH>75%)No-load operation,After stead state 30 min. | ||||||||
Temp.humi resolution | 0.01°C; 0.1%RH | ||||||||
20°C~High TempHeating up time | °C | 100 150 | |||||||
Min | 30 40 | 30 40 | 30 45 | 30 45 | 30 45 | 30 45 | |||
20°C~Low tempCooling down time | °C | 0 -20 -40 -60 -70 | |||||||
Min | 25 40 50 70 80 | ||||||||
Heating rate | ≥3°C/min | ||||||||
Cooling rate | ≥1°C/min | ||||||||
Material
|
Inner chamber material | SUS#304 stainless steel plate | |||||||
Exterior chamber material | Stainless steel plate+ powder coated | ||||||||
Insulation Material | PU & Fiberglass wool | ||||||||
System
|
Air circulation system | Cooling fan | |||||||
Fan | Sirocco fan | ||||||||
Heating System | SUS#304 stainless steel high-speed heater | ||||||||
Air flow | Forced Air Circulation(It enters at bottom and leaves at top) | ||||||||
Humidification system | Surface evaporation system | ||||||||
Refrigeration system | Imported compressor,French Tecumseh compressor or German Bitzer Compressor,finned type evaporator,air(Water)-cooling condenser | ||||||||
Refrigerating fluid | R23/ R404A USA Honeywell. | ||||||||
Condensation | Air(Water)-cooling condenser | ||||||||
Dehumidifying system | ADP critical dew point cooling/dehumidifying method | ||||||||
Controlling system | Digital electronic indicators+S.S.R.With PID automatic calculation capability | ||||||||
Operation interface | Grande Expertise in Temperature & Humidity Controller,Chinese-English Shift. | ||||||||
Controller
|
Programmable capability | Save 120 profiles with up to 1200 steps each | |||||||
Setting range | Temperature:-100℃+300℃ | ||||||||
Reading accuracy | Temperature:0.01℃ | ||||||||
Input | PT100 or T Sensor | ||||||||
Control | PID control | ||||||||
Communication interface | Equipped with standard communication interface devices USB,RS-232 and RS-485,enable the test chamber to be connected with personal computer(PC),to achieve multi-machine control and management at the same time.Standard: USB external memory port.Optional: RS-232, RS-485, GP-IB, Ethernet | ||||||||
Print function | Japan Yokogawa Temperature Recorder(Optional accessories) | ||||||||
Auxiliary | Limit Alarm,Self Diagnosis,Alarm Display(Failure Cause),Timing device(Automatic Switch) | ||||||||
Accessories | Multi-layer vacuum glass observation window,Cable port(50mm),Controlling status indicator lamp,Chamber light,Specimen loading shelf(2pcs,position adjustable),Guaze 5pcs,Operation manual 1 set. | ||||||||
Safety protection device | Over-heat protection circuit breaker,Compressor overload protection,Control system overload protection,Humidifying system overload protection,Overload indicator lamp. | ||||||||
Power supply | AC 1Ψ 110V;AC 1Ψ 220V;3Ψ380V 60/50Hz | ||||||||
Customized service | Welcome to Non-standard,Specail requirements,OEM/ODM orders. | ||||||||
The technical information will be subjected to change without notice |
● High performance and quiet operation (65 dBa)
● Space-saving footprint, designed for flush installation to wall
● Stainless steel exterior
● Full thermal break around doorframe
● One 50mm (2") or 100mm (4") diameter cable port on left, with flexible silicone plug
● Three levels of overheat protection, plus overcool protection
● Easy lift-off service panels, electrical access on left
● Detachable eight-foot power cord with plug
● ETL listed electrical panel conforming to UL 508A
Touch-screen programmer/controller with Ethernet
Save 120 profiles with up to 1200 steps each (ramp, soak, jump, auto-start, end)
One event relay for external device control, plus specimen power interlock relay for safety
Grande exclusive options include: Web Controller for full remote access; Chamber Connect software for basic data logging and monitoring. USB and RS-232 ports available, as well.
● GB11158 high-temperature testing condition
● GB10589-89 low-temperature testing condition
● GB10592-89 high-low-temperature testing condition
● GB/T10586-89 humidity testing condition
● GB/T2423.1-2001 low-temperature testing condition
● GB/T2423.2-2001 high-temperature testing condition
● GB/T2423.3-93 humidity testing condition
● GB/T2423.4-93 alternating temperature testing machine
● GB/T2423.22-2001 temperature testing method
● EC60068-2-1.1990 low-temperature testing method
● IEC60068-2-2.1974 high-temperature testing method
● GJB150.3 high-temperature test
● GJB150.3 high-temperature test
● GJB150.9 humidity test